Medical Device Recall: AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
Ad-Tech Medical Instrument Corporation · September 19, 2024
Reason for Recall
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Distribution
US Nationwide distribution in the states of CA, OH.
States Affected
NATIONWIDE
Quantity Affected
4 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0220-2025
Status: ongoing