Ad-Tech Medical Instrument Corporation Recalls
3 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (3)
- Class II
September 19, 2024
AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
- Class II
September 19, 2024
AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
- Class II
September 19, 2024
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.