Medical Device Recall: BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris
CareFusion 303, Inc. · October 17, 2025
Reason for Recall
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Product Description
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.
States Affected
NATIONWIDE
Quantity Affected
2,574,396
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0430-2026
Status: ongoing