CareFusion 303, Inc. Recalls
84 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (84)
- Class II
February 11, 2026
BD Alaris" System with Guardrails" Suite MX with Point of Care Unit
Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
- Class II
December 3, 2025
BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136
Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.
- Class II
November 20, 2025
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- D
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
- Class II
November 6, 2025
BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent a
Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network
- Class I
October 17, 2025
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
- Class II
September 30, 2025
BD Pyxis Pro MedStation Main, REF: 1155-00
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
September 30, 2025
Various models of BD Pyxis MedFlex, Reference numbers: 1119-00 1139-00 139038-01 139039-01 139040-01 139041-01 139043-01 139044-01 139045-01 139046-01 139049-01 139051-01 139052-01 13905
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
September 30, 2025
Various models of BD Pyxis Medbank: Reference numbers: 1137-00 1145-00 1146-00 1147-00 1148-00 138902-01 138903-01 138905-01 138906-01 138907-01 138908-01 138909-01 138910-01 138911-01
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
September 30, 2025
All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
September 30, 2025
Various models of BD Pyxis Pro 7-Drawer Auxiliary, Reference numbers: 1149-00 and 1152-00
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class I
September 11, 2025
BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve S
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Class III
July 10, 2025
Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS MEDBANK MINI 3HH-1HM/169-111, BD PYXIS MEDBANK MINI 3HH-1
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
- Class I
July 8, 2025
BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Class I
July 8, 2025
BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV; REF: 2452-0007, BD Alaris Pump Infusion Set, 2 Back Check
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Class II
May 8, 2025
Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main 10885403512629 [303], BD Pyxis MedStation ES Tower 10885403512674 [352]
During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.
- Class II
May 8, 2025
BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower / 10885403512674 / 352 BD Pyxis" Anesthesia Station ES /
Antivirus software was not consistently installed on impacted devices during the implementation process.
- Class II
March 11, 2025
BD PYXIS MEDBANK (Name, REF): MEDPASS MODULE, 139088-01; MINI 1FH-1FM, 169-137; MINI 1HH-1FH-1HM, 169-101; MINI 1HH-1HM-1FM, 169-102; MINI 2FH, 138902-01; MINI 2FH, 169-104; MINI 2FM, 169-105;
Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.
- Class II
March 11, 2025
BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qua
Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.
- Class II
March 11, 2025
BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-
Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients.
- Class I
February 18, 2025
BD Care Coordination Engine (CCE) Infusion Adapter, Safety Management Software
Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
- Class I
February 18, 2025
BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software
Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
- Class II
February 6, 2025
BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
- Class II
January 23, 2025
BD Pyxis MedFlex 2.0 REF 1139-00 Bio-ID Spare Parts REF 353200-01 The BD Pyxis" MedFlex is an Automated Dispensing Cabinet (ADC) intended to securely store and dispense medications and supplies
Due to fingerprint scanner failing resulting in the scanner heating up.
- Class II
January 23, 2025
BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, p
Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
- Class II
January 10, 2025
BD Pyxis: MedFlex, REF: 1119-00; MedBank Tower REF: 1145-00; MedBank MedPass, REF: 139088-01 and associated BD Pyxis user/software/safety guides and quick reference guides.
Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.
- Class II
January 8, 2025
139088-01 BD PYXIS MEDBANK MEDPASS MODULE 169-151 BD PYXIS MEDBANK MINI 1FH-2HH MEDPASS 169-152 BD PYXIS MEDBANK MINI 2HH-1FM MEDPASS 169-150 BD PYXIS MEDBANK MINI 4HH MEDPASS 169-140 BD PYXIS MEDBANK
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- Class II
January 8, 2025
139054-01 BD PYXIS MEDFLEX MN 1000 10HH-1FM 139055-01 BD PYXIS MEDFLEX MN 1000 12HH 139053-01 BD PYXIS MEDFLEX MN 1000 2HH-2FH-3FM 139041-01 BD PYXIS MEDFLEX MN 1000 2HH-2FM 139046-01 BD PYXIS MEDFLEX
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- Class II
January 8, 2025
138913-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138911-01 BD PYXIS MEDBANK MINI CR-4HH-P 169-115 BD PYXIS MEDBANK MINI CR-4HH-P 139001-01 BD PYXIS MEDBANK MN 20
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- Class II
January 8, 2025
(1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis MedStation 4000 Console, Catalog
Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.
- Class II
January 8, 2025
155288-01 BD PYXIS MEDBANK MINI 1FH-1FM 169-137 BD PYXIS MEDBANK MINI 1FH-1FM 155311-01 BD PYXIS MEDBANK MINI 1FH-2HH 138921-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 155159-01 BD PYXIS MEDBANK MINI 1HH-1F
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- Class II
January 8, 2025
(1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis CII Safe ES, Catalog No. 1116-00
Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues
- Class II
November 14, 2024
BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- Class II
November 14, 2024
BD Pyxis CII Safe ES, REF: 1116-00
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- Class II
November 14, 2024
BD Pyxis MedStation ES, REF: 323
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- Class II
August 15, 2024
Alaris System PC Unit Model 8015 - a modular infusion pump and monitoring system
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
- Class II
August 15, 2024
Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
- Class II
August 15, 2024
Alaris Syringe Module Model 8110, a modular infusion pump and monitoring system
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
- Class II
August 15, 2024
Alaris PCA Module Model 8120, a modular infusion pump and monitoring system
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
- Class II
August 15, 2024
Alaris Pump Module Model 8100, a modular infusion pump and monitoring system
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
- Class II
July 9, 2024
BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system
Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.
- Class II
July 9, 2024
BKIT BD Pyxis QFill Replenishment Station, REF: 155197-01, medication cabinet system
Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P, REF: 169-82 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-12HH-1FH, REF: 169-84 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P, REF: 169-97 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P, REF: 169-91 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P, REF: 169-90 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-14HH-P, REF: 169-86 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK MINI CR-4HH-P, REF: 169-115 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P, REF: 169-93 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P, REF: 169-99 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P, REF: 169-94 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow, REF: 169-131 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-2HH-6FH, REF: 169-88 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK MINI CR-2HH-1FH-P, REF: 169-114 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P, REF: 169-92 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P, REF: 169-98 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P, REF: 169-85 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P, REF: 169-83 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
June 24, 2024
BD PYXIS MEDBANK TWR MN CR-2HH-6FM, REF: 169-89 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
- Class II
May 21, 2024
BD Alaris System Manager REF: 9601, compatible with BD Alaris PC Units
Due to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server.
- Class II
May 13, 2024
BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1HH-1HM-1FM/ 169-102, MEDBANK MINI 1HH-3HM/ 169-103, ME
Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.
- Class II
May 13, 2024
BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-
Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.
- Class II
May 2, 2024
BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII S
If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.
- Class II
April 15, 2024
BD Pyxis" MedStation ES 7-Drawer Auxiliary Tower, REF: 324
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
April 15, 2024
BD Pyxis MedStation ES, REF: 323, Medication Cabinet
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
April 15, 2024
BD Pyxis(TM) Anesthesia Station 4000, REF: 338
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
April 15, 2024
BD Pyxis" Anesthesia Station ES, REF: 327
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
April 15, 2024
BD Pyxis MedStation 4000 System, REF: 303, Medication Cabinet
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
April 15, 2024
BD Pyxis MedStation 4000 7-Drawer Auxiliary Tower, REF: 306
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
- Class II
February 23, 2024
BD Pyxis MedStation Auxiliary 4000 REF: 306. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
- Class II
February 23, 2024
BD Pyxis MedStation Auxiliary ES REF: 324. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
- Class II
February 23, 2024
BD Pyxis MedStation ES (Med ES Main) REF: 323. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
- Class II
February 23, 2024
BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
- Class II
February 1, 2024
BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;
Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.
- Class II
October 5, 2023
MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD P
If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.
- Class II
April 5, 2023
BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.
- Class II
February 14, 2023
BD Pyxis MedStation ES AUX, REF 324
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Class II
February 14, 2023
BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" MedStation ES and Pyxis" MedStation 4000 are automated dispensing cabinets (ADC) that are inte
Due to an increase in complaints of door latch assembly failures that may cause the doors to be left in an open state potentially causing inventory discrepancies, a delay in access to medication and/or diversion of medication.
- Class II
February 14, 2023
BD Pyxis MedStation 4000 Main, REF 303
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Class II
February 14, 2023
BD Pyxis Med 4000 Auxiliary (AUX), REF 314
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Class II
February 14, 2023
BD Pyxis Med ES Auxiliary (AUX), REF 343
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Class II
February 14, 2023
BD Pyxis MedStation ES (Med ES Main), REF 323
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Class II
February 14, 2023
BD Pyxis MedStation 4000 AUX, REF 306
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,