Medical Device Recall: BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qua
CareFusion 303, Inc. · March 11, 2025
Reason for Recall
Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.
Product Description
BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.
Distribution
US Nationwide distribution: CA, TX, WI, CO, AK, VA, TN, WA, NC, NY, MS, FL, KS, MI, IN, MN, MT, KY, MO, UT, NJ, OH, PA, ND, OR, NE, SD, NV, OK, GA, MA, IA, HI, ME, AZ, IL, SC, LA, WV, AL, ID, NM, DC, NH, VT, CT, MD, AR, WY.
States Affected
NATIONWIDE
Quantity Affected
7357
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1545-2025
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.