Class II
Medical Device Recall: STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test
Diagnostica Stago, Inc. · July 29, 2026
Reason for Recall
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
Distribution
US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.
States Affected
NATIONWIDE
Quantity Affected
5220 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-3047-2026
Status: ongoing