- Class II
February 5, 2026
Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: The STA¿ - Liatest ¿ D-Di kit is an immuno-turbidimetric assay for the quantitative
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
- Class II
February 4, 2026
Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use o
The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
- Class II
December 15, 2025
Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generat
Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
- Class II
September 12, 2025
Product: STA Liatest D-Di; REF: 00515;
A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
- Class II
July 21, 2025
STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of th
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Class II
July 21, 2025
STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromogenic, and immunolo
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Class II
July 21, 2025
STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determi
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Class II
July 21, 2025
STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of t
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Class II
May 21, 2025
STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%