Class II

Medical Device Recall: FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routin

FUJIFILM Healthcare Americas Corporation · April 11, 2024

Reason for Recall

Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Product Description

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Distribution

US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.

States Affected

NATIONWIDE

Quantity Affected

18 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1898-2024

Status: ongoing