FUJIFILM Healthcare Americas Corporation Recalls
15 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (15)
- Class II
January 9, 2026
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
- Class II
September 15, 2025
FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.
- Class II
June 11, 2025
ASPRIRE Cristalle Mammography System (cleared under K212873) installed with the ASPIRE Cristalle Digital Breast Tomosynthesis (DBT) Option (approved under P160031).
Devices had an unapproved slabbing software function enabled for use.
- Class II
February 11, 2025
Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
The incorrect computed patient age is showing in VX for patients less than 3 months old.
- Class II
April 11, 2024
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
- Class II
April 11, 2024
FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routin
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
- Class II
January 10, 2024
Synapse PACS - Version 7.3.000
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 10, 2024
Synapse PACS - Version 7.2.200
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 10, 2024
Synapse PACS - Version 7.2.100
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 10, 2024
Synapse PACS - Version 7.1.000
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 10, 2024
Synapse PACS - Version 7.2.000
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 10, 2024
Synapse PACS - Version 7.1.000US
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
- Class II
January 3, 2024
Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
- Class II
January 3, 2024
Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
- Class II
November 23, 2023
The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.