Medical Device Recall: Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

FUJIFILM Healthcare Americas Corporation · April 11, 2024

Reason for Recall

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

418 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1884-2024

Status: ongoing