Class II

Medical Device Recall: Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

FUJIFILM Healthcare Americas Corporation · January 3, 2024

Reason for Recall

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Distribution

US domestic distribution nationwide. International distribution to Israel.

States Affected

NATIONWIDE

Quantity Affected

13

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1171-2024

Status: ongoing