Class II

Medical Device Recall: The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

FUJIFILM Healthcare Americas Corporation · November 23, 2023

Reason for Recall

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

Distribution

US Nationwide and Worldwide

States Affected

NATIONWIDE

Quantity Affected

130

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3158-2024

Status: ongoing