Class II

Medical Device Recall: Allia Moveo angiographic X-ray system

GE Medical Systems, LLC · April 20, 2026

Reason for Recall

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Distribution

Worldwide - US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

3 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2318-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.