GE Medical Systems, LLC Recalls
98 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (98)
- Class II
March 26, 2026
Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution Apex Elite, X-ray Computed Tomography, Model Number 5590000-23
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
March 26, 2026
Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Class II
February 3, 2026
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
- Class II
January 30, 2026
Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-02
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
- Class II
January 30, 2026
Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
- Class II
January 30, 2026
Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 58
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
- Class II
October 24, 2025
Senographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and diagnostic mammography
Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR ¿1020.30(j).
- Class II
September 18, 2025
LOGIQ P10 series with software version R4.5.7 Model Number 5877534
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
- Class II
September 18, 2025
LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
- Class II
September 18, 2025
LOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
- Class II
August 22, 2025
MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
- Class II
July 2, 2025
GE HealthCare OEC 9900 Elite C-arm Systems.
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
- Class II
July 2, 2025
GE HealthCare OEC 9800.
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
- Class II
June 20, 2025
GE SIGNA Premier MAX (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Discovery MR750 3.0T, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Discovery MR450 1.5T, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Hero XT (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Premier, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Artist Evo, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA PET/MR AIR (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA UHP, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Optima MR450w 1.5T, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Voyager AIR (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA PET/MR AIR Plus (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Signa Artist, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Premier Elite (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Hero, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Architect, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Discovery MR750w 3.0T, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Premier XT (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE Signa Voyager, Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography with Nuclear Magnetic Resonance
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Hero Elite (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Hero MAX (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
June 20, 2025
GE SIGNA Architect AIR (China Only), Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
- Class II
May 23, 2025
SIGNA PET/MR
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Class II
May 23, 2025
Discovery MR750w 3.0T
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Class II
May 23, 2025
SIGNA Architect
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Class II
May 23, 2025
SIGNA Architect AIR
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Class II
March 21, 2025
OEC 3D
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Class II
March 21, 2025
OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, intervent
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Class II
March 6, 2025
Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP f
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
- Class II
March 6, 2025
Venue Fit Models: R3 (Model Number H45303VFSC, H45303VFSW), R4 (Model Number H45314VFS, H45314VFSW), Venue Fit R5 system (Model Numbers H45335VFS, H45335VFSW), R3 Surgery for China, R3 PeriOP for Chin
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
- Class II
February 18, 2025
GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
February 18, 2025
GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Class II
January 28, 2025
Discovery XR656HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Class II
January 28, 2025
Optima XR646 HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Class II
January 28, 2025
Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Class II
January 28, 2025
Discovery XR656 HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Class II
December 6, 2024
Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facil
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
- Class II
November 25, 2024
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8)
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
- Class II
November 25, 2024
GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
- Class II
November 25, 2024
GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
- Class II
November 18, 2024
Tec 820 ISO Anesthesia Vaporizer, Reference Number 1177-9820-ISO
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
- Class II
November 18, 2024
Tec 820 SEV Anesthesia Vaporizer, Reference Number 1177-9820-SEV
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
- Class II
November 18, 2024
Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 2066835-001, 2069767-001, 207
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
- Class II
November 18, 2024
Tec 850 ISO Anesthesia Vaporizer, Reference Number 1177-9850-ISO
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
- Class II
November 18, 2024
Tec 850 SEV Anesthesia Vaporizer, Reference Number 1177-9850-SEV
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
- Class II
October 28, 2024
Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Class II
October 28, 2024
SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Class II
October 28, 2024
Discovery MR750w 3.0T, whole body magnetic resonance scanner
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Class II
October 28, 2024
SIGNA Architect, Nuclear Magnetic Resonance Imaging System
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Class II
June 14, 2024
Discovery MR450 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Voyager, with affected software versions: PX26.0 to PX26.6, VX28.0, VX29.1, VX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Optima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Creator, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA MR355, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Discovery MR750w 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.5, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Pioneer, with affected software versions: PX25.0 to PX25.5, PX26.0, PX26.1, PX28.0 to PX28.4, PX29.0, PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Artist, with affected software versions: DV26.0, DV27.1, DV28.1, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Brivo MR355, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Prime, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA MAGNUS, with affected software versions: MR29.1, RX29.1; Nuclear Magnetic Resonance Imaging System - research use only
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Premier, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA UHP, with affected software versions: RX28.0, MR30.1; Nuclear Magnetic Resonance Imaging System - research use only
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA MR380, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA MR360, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Discovery MR750 3.0T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0 to DV26.4, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA 7.0T, with affected software versions: 7T29.1, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
3.0T SIGNA HDxt Family, with affected software versions: RX27.0 to RX27.3, RX28.0, RX29.0 to RX29.2, MR30.0, MR30.1HD16.0_V03, HD23.0_V03; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
Discovery MR950, with affected software versions: 7T23.0; Nuclear Magnetic Resonance Imaging System - research use only
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Class II
June 14, 2024
SIGNA Hero, with affected software versions: PX29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.