Class II

Medical Device Recall: GE HealthCare OEC 9900 Elite C-arm Systems.

GE Medical Systems, LLC · July 2, 2025

Reason for Recall

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2342-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.