Medical Device Recall: GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB

GE Medical Systems, LLC · November 25, 2024

Reason for Recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

Distribution

Worldwide

Quantity Affected

53 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0811-2025

Status: ongoing