Medical Device Recall: GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
GE Medical Systems, LLC · February 18, 2025
Reason for Recall
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Distribution
Worldwide distribution.
Quantity Affected
21 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1303-2025
Status: ongoing