Medical Device Recall: MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4

GE Medical Systems, LLC · August 22, 2025

Reason for Recall

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom.

States Affected

NATIONWIDE

Quantity Affected

29,225 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0231-2026

Status: ongoing