Medical Device Recall: MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
GE Medical Systems, LLC · August 22, 2025
Reason for Recall
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
States Affected
NATIONWIDE
Quantity Affected
29,225 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0231-2026
Status: ongoing