Medical Device Recall: ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
Globus Medical, Inc. · December 15, 2023
Reason for Recall
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Distribution
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.
States Affected
NATIONWIDE
Quantity Affected
570
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1007-2024
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.