- Class II
April 3, 2025
MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
- Class II
December 15, 2023
ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
- Class II
December 11, 2023
Globus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref : 6143.1834S
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
- Class II
December 11, 2023
Globus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1842S
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
- Class II
December 11, 2023
Globus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1870S
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
- Class I
December 7, 2023
ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100
A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.