Class II

Medical Device Recall: DUR PCA MTK REV INS LFT Intended for knee replacement Product Number: 6728-2-609

Howmedica Osteonics Corp. · December 1, 2023

Reason for Recall

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Distribution

Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

States Affected

NATIONWIDE

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0919-2024

Status: ongoing