Howmedica Osteonics Corp. Recalls
91 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (91)
- Class II
February 2, 2026
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
- Class II
November 12, 2025
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit fo
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
- Class II
August 28, 2025
1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;
a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
- Class II
September 4, 2024
Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee S
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
- Class II
August 7, 2024
MAKO Integrated Cutting System (MICS) HANDPIECE, Catalog Number 209063
Stryker has identified an issue with the characterization process that impacted the constant values assigned to the associated MICS Handpieces.
- Class II
August 5, 2024
Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
A patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.
- Class II
June 19, 2024
UNITRAX Endoprosthesis Head Component - 38mm. Implantable component used in partial hip arthroplasty. Catalog Number: 6942-5-038
Mislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
- Class II
June 19, 2024
UNITRAX Endoprosthesis Head Component - 43mm. Implantable component used in partial hip arthroplasty. Catalog Number: 6942-5-043
Mislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
- Class II
June 19, 2024
UNITRAX Endoprosthesis Head Component - 42mm. Implantable component used in partial hip arthroplasty. Catalog Number: 6942-5-042
Mislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
- Class II
May 23, 2024
Gray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
- Class II
May 23, 2024
HRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER48D-indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-48D
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER54E- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-54E
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII PSL CLUSTER64H- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 742-11-64H
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER50D- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-50D
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII PSL CLUSTER58F-Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 742-11-58F
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER58F- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-58F
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII PSL CLUSTER52E- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 742-11-52E
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER66H- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-66H
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER44B- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-44B
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII PSL CLUSTER48D- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 742-11-48D
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII HEMI CLUSTER52E-Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 702-11-52E
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 27, 2024
TRIDENTII PSL CLUSTER50D- Indicated for use in primary and revision patients at high risk of hip dislocation Catalog Number: 742-11-50D
The acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
- Class II
February 6, 2024
NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
X3 TRIATHLON CS INSERT NO 6 11 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
NO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
X3 TRIATHLON CS INSERT NO 7 9 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
X3 TRIATHLON CS INSERT NO 6 10 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
February 6, 2024
X3 TRIATHLON CS INSERT NO 3 10 MM
Potential packaging breaches of inner blister and outer sterile blister.
- Class II
December 22, 2023
HRIS ACET CUP CUT TIP 32X140 Part Number: 6210-5-200rker. single-use, sterile surgical instruments used during hip arthroplasty
Failed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility
- Class II
December 22, 2023
HRIS ACET CUP CUT TIP 26X140 Part Number: 6210-5-100. single-use, sterile surgical instruments used during hip arthroplasty
Failed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7524
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0910
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32J
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement Product Number: 6637-0-831
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
TIBIAL BEARING INSERT SERIES P-S I ASSY Intended for knee replacement Product Number: 7291-0324
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-0910
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-0110
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee replacement Product Number: 6637-4-231
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacment Part Number: 72-4-7516
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0908
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DURATION A-P LIPPED TIB.INSERT-DURAC Intended for knee replacement Product Number: 6642-1-709
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #5 -Intended for knee replacement Part Number: 73-0510
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32J
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT. Intended for knee replacement Part Number: 72-4-7518
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LEFT LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-231
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7514
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3910
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0910
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
PRIMARY REV.TIB.INSERT-DURACON Intended for knee replacement. Product Number: 6637-4-031
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-411
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LEFT SMALL-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-631
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DURATION A-P LIPPED TIB.INSERT-DURAC Intended for knee replacement Product Number: 6642-1-911
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0028
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0321
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LEFT SMALL-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-028
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0030
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DUR PCA MTK REV INS LFT Intended for knee replacement Product Number: 6728-2-609
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DUR PCA MTK REV INS LFT Intended for knee replacement Product Number: 6728-2-611
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-413
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DUR PCA MTK REV INS RT Intended for knee replacement Product Number: 6728-2-711
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-28J
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7521
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
P7 28MM 10 DEGREE +4MM INSERT Intended for knee replacement Part Number: 6302-6-107
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0524
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-931
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LEFT LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-228
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- Intended for knee replacement Part Number: 72-15-0324
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
LEFT X-LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-328
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
P7 32MM 1O DEGREE +4MM INSERT Intended for knee replacement Part Number: 6302-6-307
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-2110
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
SCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DURATION PLASTIC PATELLA-DURACON Intended for knee replacement Product Number: 6642-2-200
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement Product Number: 6637-0-928
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Class II
December 1, 2023
DUR PCA MTK REV INS RT Intended for knee replacement Product Number: 6728-2-709
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.