Class II

Medical Device Recall: DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER

Howmedica Osteonics Corp. · July 30, 2026

Reason for Recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Product Description

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER - GLENOID #11 5361-6111 SHOULDER - GLENOID #5 5361-6105 SHOULDER - GLENOID #7 5361-6107 SHOULDER - GLENOID #9 5361-6109 SHOULDER - KEELED GLENOID #11 5361-6211 SHOULDER - KEELED GLENOID #5 5361-6205 SHOULDER - KEELED GLENOID #7 5361-6207 SHOULDER - KEELED GLENOID #9 5361-6209 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

States Affected

NATIONWIDE

Quantity Affected

12,156 US

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3200-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.