Medical Device Recall: IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Philips North America · August 30, 2024
Reason for Recall
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
Distribution
US Nationwide. Global Distribution.
States Affected
NATIONWIDE
Quantity Affected
545 devices
Source: FDA (openFDA Device Enforcement)
Recall number: Z-3249-2024
Status: ongoing