Philips North America Llc Recalls
100 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (100)
- Class II
August 21, 2026
Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Ingenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3. 781396(1.5T), 4. 782101(1.5T), 5. 782115(1.5T), 6. 782140(1.5T), 7. 781261(1.5T CX), 8. 781
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
MR 5300 Model Number (REF): 1. 782110, 2. 782152;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Inte
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Achieva XR; Model Numbers (REF): 1. 781153, 2. 781253;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Ingenia Elition S and Ingenia Elition X; Model Numbers (REF): 1. 781357(S), 2. 782106(S), 3. 782137(S), 4. 781358(X), 5. 782107(X), 6. 782119(X), 7. 782136(X), 8. 782158(X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 21, 2026
Ingenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4. 781271(3.0T CX), 5. 782105(3.0T CX);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- Class II
August 7, 2026
Philips Brilliance CT 64: Brilliance CT 64 Channel Product Code: 728231, Brilliance CT 64 Upgrades Product Code: 728232. Software 4.1.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
- Class II
August 7, 2026
Philips Ingenuity CT Product Codes: Ingenuity CT Upgrades 728327, Ingenuity CT China 728324, Ingenuity Core 128 Elite Brazil 728325. Software 4.4.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
- Class II
August 7, 2026
Philips Brilliance. Models: Brilliance iCT, Brilliance iCT Upgrades, Brilliance iCT SP. Software 4.1.10.x Product Codes: 728306, 728305, 728311.
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
- Class II
August 7, 2026
Philips IQon Spectral CT Product Codes: IQon Spectral CT 728332, IQon Spectral CT 728331 Software 4.7.7.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
- Class II
July 31, 2026
Philips PIC iX Essentials REF:867093
Due to software malfunction, patient information system may freeze and become unresponsive.
- Class II
July 31, 2026
Philips Patient Information Center iX REF:866389
Due to software malfunction, patient information system may freeze and become unresponsive.
- Class II
July 1, 2026
Cardiac Workstation 5000. Product Number: 860439.
Device will intermittently provide inaccurate data related to native heart rate.
- Class II
July 1, 2026
Cardiac Workstation 7000. Product Number: 860441.
Device will intermittently provide inaccurate data related to native heart rate.
- Class II
June 5, 2026
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
- Class II
May 26, 2026
Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
- Class II
May 26, 2026
Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
- Class II
May 1, 2026
DigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.
System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.
- Class II
April 14, 2026
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458,
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K2
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k)
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips SmartPath to dStream for 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781260. 2. Model Number (REF): 782112.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063,
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
April 14, 2026
Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Class II
January 7, 2026
IQon Spectral CT; Product Code (REF): 728332;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
- Class II
January 7, 2026
Spectral CT 7500 on Rails; Product Code (REF): 728334;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
- Class II
January 7, 2026
Brilliance iCT; Product Code (REF): 728306;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
- Class II
January 7, 2026
Spectral CT; Product Code (REF): 728333;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
- Class II
December 22, 2025
Patient Information Center iX; Software Version Number: 4.5.0
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
- Class II
December 12, 2025
Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
- Class II
December 3, 2025
Ingenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Evolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Ingenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Evolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
December 3, 2025
Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Class II
October 31, 2025
IntelliVue Patient Monitor MX800. Product Number: 865240.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX500. Product Number: 866064.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP50. Product Number: M8004A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX550. Product Number: 866066.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Multi-Measurement Module X3. Product Number: 867030.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP2. Product Number: M8102A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX750. Product Number: 866471.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX450. Product Number: 866062.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX400. Product Number: 866060.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX700. Product Number: 865241.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP90. Product Number: M8010A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP70. Product Number: M8007A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX600. Product Number: 865242.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX850. Product Number: 866470.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP60. Product Number: M8005A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP30. Product Number: M8002A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP80. Product Number: M8008A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Patient Monitor MX100. Product Number: 867033.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP40. Product Number: M8003A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue Multi Measurement Server X2. Product Number: M3002A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP5. Product Number: M8105A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 31, 2025
IntelliVue MP20. Product Number: M8001A.
Potential issue where the IntelliVue monitors did not alarm.
- Class II
October 29, 2025
CT 5300; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
- Class II
October 29, 2025
Incisive CT; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
- Class II
October 27, 2025
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
- Class II
September 12, 2025
Cardiac Workstation 7000; Model Number: 860441;
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
- Class II
September 12, 2025
Cardiac Workstation 5000; Model Number: 860439;
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Brilliance 16 Power, Model Number: 728240. 2. Brilliance CT 16 Slice, Model Number: 728246. 3. Brilliance CT 6 Slice, Model Number: 728256.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Brilliance iCT SP, Model Number: 728311.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326, Model Number: 728323. 3. Ingenuity CT, Model Number: 728326.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2. Brilliance CT Big Bore Oncology, Model Number: 728243. 3. Brilliance CT Big Bore, Model Number: 728244.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
August 13, 2025
Philips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Class II
June 30, 2025
Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.