Medical Device Recall: Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
Philips North America Llc · May 3, 2024
Reason for Recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
States Affected
NATIONWIDE
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1844-2024
Status: ongoing