Class II

Medical Device Recall: Ingenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3. 781396(1.5T), 4. 782101(1.5T), 5. 782115(1.5T), 6. 782140(1.5T), 7. 781261(1.5T CX), 8. 781

Philips North America Llc · August 21, 2026

Reason for Recall

A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

Product Description

Ingenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3. 781396(1.5T), 4. 782101(1.5T), 5. 782115(1.5T), 6. 782140(1.5T), 7. 781261(1.5T CX), 8. 781262(1.5T CX), 9. 781347(1.5T S);

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;

States Affected

NATIONWIDE

Quantity Affected

2038

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3236-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.