Class II

Medical Device Recall: IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows: 1) IntelliVue MX400 866060; 2) IntelliVue MX450 866062;

Philips North America Llc · June 14, 2023

Reason for Recall

Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.

Product Description

IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows: 1) IntelliVue MX400 866060; 2) IntelliVue MX450 866062; 3) IntelliVue MX500 866064; 4) IntelliVue MX550 866066 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

Distribution

Foreign Only: Denmark France Germany Japan United Kingdom

Quantity Affected

153 units OUS

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0020-2025

Status: ongoing