Medical Device Recall: Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider E
Philips North America Llc · March 22, 2024
Reason for Recall
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
Product Description
Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model: SCL500RMI1U
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, United Kingdom, Utd. Ara Emir.
States Affected
NATIONWIDE
Quantity Affected
12,936 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1690-2024
Status: ongoing