Medical Device Recall: Zenition 70. Product Code 718133. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to
Philips North America Llc · February 25, 2025
Reason for Recall
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Product Description
Zenition 70. Product Code 718133. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the devices.
Distribution
US
Quantity Affected
N/A
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1581-2025
Status: ongoing