Class II
Medical Device Recall: Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
Pivotal Health Solutions, Inc. · October 9, 2025
Reason for Recall
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Distribution
Distribution to US nationwide, Canada and Japan.
States Affected
NATIONWIDE
Quantity Affected
464 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0588-2026
Status: ongoing