Pivotal Health Solutions, Inc. Recalls
3 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (3)
- Class II
October 9, 2025
Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
- Class II
October 9, 2025
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
- Class II
April 29, 2024
Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitioner definable target treatment location, treatment pounds, and patterns of treatment.
This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.