Class II

Medical Device Recall: Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute is

Scientia Vascular, Inc. · January 2, 2025

Reason for Recall

Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

Product Description

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Distribution

U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.

States Affected

NATIONWIDE

Quantity Affected

120 devices

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1234-2025

Status: ongoing