Scientia Vascular, Inc. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
January 16, 2025
Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during inte
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
- Class II
January 2, 2025
Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute is
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.