Medical Device Recall: 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not
Stryker Corporation · September 5, 2024
Reason for Recall
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
Product Description
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
States Affected
NATIONWIDE
Quantity Affected
307 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0063-2025
Status: ongoing