Stryker Corporation Recalls
57 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (57)
- Class II
April 9, 2026
Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are intended to be used with the FluidSmart Fluid Management System, which is intended for irri
Due to nonconforming products being inadvertently distributed.
- Class II
January 21, 2026
MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009 (Pack of 10) - MOLLI Introducer (12cm, 14G) and MOLLI Marke
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
- Class II
October 31, 2025
NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
- Class II
October 31, 2025
NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 Bra
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
- Class II
October 31, 2025
NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP; NN-8003 13x10 MHP; NN-8004 13x13 MHP; NN-8005 15x25 MHP GAAB; N
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
- Class II
July 2, 2025
Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
- Class II
June 18, 2025
Stryker, StrykeFlow REF: 0250070520 / 2, Disposable Suction/Irrigator with Disposable Tip
Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.
- Class II
June 18, 2025
Stryker, StrykeFlow , REF: 0250070500, 2, Disposable Suction/Irrigator without Disposable Tip
Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.
- Class II
May 28, 2025
Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
- Class II
May 22, 2025
Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
- Class II
May 22, 2025
Femoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Number: 1851-0830S.
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
- Class II
March 28, 2025
Guide Wire with Ruler Tube 3x800 mm DIA; Catalog number/REF: 2351-3080S;
The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.
- Class II
March 28, 2025
Guide Wire with Ruler Tube 3x1000 mm DIA; Catalog number/REF: 2351-3100S;
The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.
- Class II
February 25, 2025
Nasopore Standard 4cm fragmentable nasal dressing, Catalog Numbers 5400-010-004 and 5400-010-004ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Otopore Square outer ear wound dressing, Firm Catalog Number 5400-020-100 and Forte Catalog Number 5400-020-100ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Otopore Cylinder Standard outer ear dressings, Catalog Numbers 5400-010-000 and 5400-010-000ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Otopore Cylinder outer ear wound dressing, Firm Catalog Number 5400-020-000 and Forte Catalog Number 5400-020-000TL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Hemopore 8cm nasal/sinus temporary wound dressing, Catalog Numbers 5400-020-208 and 5400-020-208ITL (OUS)
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Hemopore 2PK nasal/sinus temporary wound dressing, Catalog Number 5400-222-208S
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore Standard 8cm fragmentable nasal dressing, Catalog Numbers 5400-010-008 and 5400-010-008ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore 8cm fragmentable nasal dressing, Firm Catalog Number 5400-020-008 and Forte Catalog Number 5400-020-008ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and Nasopore Forte plus 4cm Catalog Number 5400-030-004ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
February 25, 2025
Nasopore 4cm fragmentable nasal dressing, Firm Catalog Number 5400-020-004 and Forte Catalog Number 5400-020-004ITL
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Class II
January 29, 2025
Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643, Sterile EO, Rx Only MR Safe CE 0197
Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.
- Class II
January 8, 2025
stryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal Blades
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
- Class II
January 8, 2025
stryker Precision Thin, Catalog Number REF 2296-003-155; Micro Oscillating and Sagittal Blades
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
- Class II
December 3, 2024
100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)
Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.
- Class II
October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke gen
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
October 4, 2024
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by el
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generate
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
October 4, 2024
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated b
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke gene
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke genera
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
September 5, 2024
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
- Class II
July 24, 2024
Stryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 3.0mm Diamond Match Head, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medi
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following me
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 4.0mm Precision Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following me
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 2.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical ap
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 5.0mm Coarse Diamond Round, Distal Bend intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following m
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 3.0mm Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical a
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following me
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
July 24, 2024
Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical ap
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
- Class II
June 3, 2024
Curved, super elastic cement delivery needle that passes through the cannula in a straight configuration and enters the vertebral body in a curved configuration. This provides physicians with the abil
Expired Products distributed to customers
- Class II
June 3, 2024
This product is a single use sterile device that is used to direct the lesion to the targeted site. Each cannula contains a stylet that is removed once the cannula is placed. The electrode is then ins
Expired Products distributed to customers
- Class II
June 3, 2024
The Non-Sterile Flyte hood covers the user s head and neck region and is intended to be worn with commercially available gowns. The hood is donned after the Flyte helmet and before the gown, such that
Expired Products distributed to customers
- Class II
May 9, 2024
Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.
The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
- Class II
December 6, 2023
Stryker Neurovascular, Trevo Trak 21 Microcatheter, REF 90338, 0.021in X162cm
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
- Class II
December 6, 2023
Stryker Neurovascular, Trevo Pro 18 Microcatheter, REF 90238 0.021in X150cm
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
- Class II
June 28, 2023
stryker 3.0MM Neuro Match Head, REF 5820-107-430, used to cut bone and bone cement during neurosurgery
There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-430 but package contains a 5820-107-530 device or vice versa).
- Class II
June 28, 2023
stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgery
There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-430 but package contains a 5820-107-530 device or vice versa).
- Class II
June 28, 2023
The 3.0MM Neuro Match Head Soft Touch bur (5820-107-030) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat
There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-030 but package contains a 5820-107-530 device).
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.