Class II

Medical Device Recall: 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

Stryker Corporation · December 3, 2024

Reason for Recall

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.

Distribution

US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.

States Affected

GA, IL, WA

Quantity Affected

14 (Expanded to 21 units on 10/9/25)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0807-2025

Status: ongoing