Class II
Medical Device Recall: Nasopore Ex Firm 4cm fragmentable nasal dressing, Catalog Number 5400-030-004 and Nasopore Forte plus 4cm Catalog Number 5400-030-004ITL
Stryker Corporation · February 25, 2025
Reason for Recall
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
182,344 total units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1478-2025
Status: ongoing