Class II

Medical Device Recall: Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235

Stryker Corporation · July 2, 2025

Reason for Recall

Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.

Distribution

Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom

States Affected

NATIONWIDE

Quantity Affected

233.230

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2277-2025

Status: ongoing