Class II

Medical Device Recall: Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cyto

Sysmex America, Inc. · January 6, 2025

Reason for Recall

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Product Description

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

Distribution

US Nationwide.

States Affected

NATIONWIDE

Quantity Affected

32 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1113-2025

Status: ongoing