Sysmex America, Inc. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
December 15, 2025
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
- Class II
January 6, 2025
Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cyto
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.