Medical Device Recalls
RSS ↗7,979 medical device recalls from federal enforcement feeds.
Showing 5,601–5,650 of 7,979
- Class II
Medical Device Recall · July 15, 2024
Product Name: 25+ CMB PAK 10K CPM,V,STD 0.9 Model/Catalog Number: 8065752450 Software Version: N/A Component: N/AAlcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
- Class II
Medical Device Recall · July 15, 2024
Product Name: 25+ TOTALPLUSr CP PAK 20K CPM BWV .9 IU Model/Catalog Number: 8065000093 Software Version: N/A Component: N/AAlcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
- Class II
Medical Device Recall · July 15, 2024
TiBrid Stand Alone Intervertebral Body Fusion Device, TiBrid SA Angled Screwdrivers, Sizes 30 degree, 45 degree, and 60 degree angles.Failure of fusion system instruments in the field.
- Class II
Medical Device Recall · July 12, 2024
GE Healthcare (China) Co., Ltd.
GE Healthcare Proteus XR/a, stationary X-Ray System, Model/Catalog Numbers: 1) 2259973; 2) 2259988; 3) 2266999; 4) 2373519; 5) 2379827; 6) 5729352; 7) 2259973-1; 8) 2259973-2; 9) 22The XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to avoid personnel injury during de-install.
- Class II
Medical Device Recall · July 12, 2024
Ion Endoluminal System, REF: 380748-65There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
- Class II
Medical Device Recall · July 12, 2024
DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrationsGround screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.
- Class I
Medical Device Recall · July 12, 2024
Amingo. OUS only.Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualifiePotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aespire 7900. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualifiePotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class II
Medical Device Recall · July 12, 2024
Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3010 Product Description: Contour Gold Eyelid Weight, 1.0 gm MedDev Contour and ThinProfile Eyelid Weight IDue to mislabeling of products
- Class I
Medical Device Recall · July 12, 2024
Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in tPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class II
Medical Device Recall · July 12, 2024
Brand Name: Contour Product Name: Contour Gold Eyelid Weight Model/Catalog Number: LL3012 Product Description: Contour Gold Eyelid Weight, 1.2 gm MedDev Contour and ThinProfile Eyelid Weight IDue to mislabeling of products
- Class III
Medical Device Recall · July 12, 2024
NavaClick Syringe and Needle, Model Number LIN02Incorrect GTIN number on carton.
- Class I
Medical Device Recall · July 12, 2024
Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, however are included duePotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
RMU-2000 Automated Chest Compression DeviceProblem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.
- Class I
Medical Device Recall · July 12, 2024
Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trainedPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualifiedPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Avance CS2 and Avance CS2 Pro. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionalsPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified iPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in thPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class I
Medical Device Recall · July 12, 2024
Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and aPotential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Class II
Medical Device Recall · July 11, 2024
2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; cMedtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
- Class II
Medical Device Recall · July 10, 2024
Ortho-Clinical Diagnostics, Inc.
The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side separated by a plastic barrier sealed within a single slide fThe reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 7600 analyzer. This dust and debris may then settle onto VITROS Na+ (Sodium) Slides as they are processed through the same MicroSlide Subsystem, potentially resulting in an increase in non-reproducible, positively, or negatively biased Na+ results. The level of bias will depend on where the dust and debris settle on the Na+ MicroSlide and may impact Na+ measurements to varying degrees inside and outside of the measuring range depending on the amount of contamination. All of which could vary from no/low bias with no patient harm, to extreme bias (falsely elevated or reduced results) potentially resulting in incorrect diagnosis; treatment; and/or a serious patient health hazard (for example, hypernatremia or hyponatremia). This issue may also potentially generate error codes preventing the analyzer from processing the MicroSlide Assays and cause delay of the patient results. All current and expired VITROS XT Chemistry Products ALB-TP slides are affected by this issue.
- Class II
Medical Device Recall · July 10, 2024
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter InstructionDue to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
- Class II
Medical Device Recall · July 10, 2024
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC)Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
- Class II
Medical Device Recall · July 10, 2024
IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiationTemporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
- Class II
Medical Device Recall · July 10, 2024
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrumentBeckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
- Class II
Medical Device Recall · July 9, 2024
Aurora 4 Series, Aurora 7 lighthead; Models: AUA7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
- Class II
Medical Device Recall · July 9, 2024
Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
- Class I
Medical Device Recall · July 9, 2024
McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx OnlyBattery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
- Class II
Medical Device Recall · July 9, 2024
DIAMOND POINT, BOTH ENDS K-WIRE, REF 128042, metallic bone fixation fastenerSmith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead of K-WIRE .045X9 2PT DM as described on the product label.
- Class II
Medical Device Recall · July 9, 2024
BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet systemDue to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.
- Class II
Medical Device Recall · July 9, 2024
Aurora 3 Series, Aurora 7 lighthead; Models: AUT7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
- Class II
Medical Device Recall · July 9, 2024
DIAMOND POINT, BOTH ENDS K-WIRE, REF 128062, metallic bone fixation fastenerSmith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead of K-WIRE .045X9 2PT DM as described on the product label.
- Class I
Medical Device Recall · July 9, 2024
McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx OnlyBattery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
- Class II
Medical Device Recall · July 9, 2024
BKIT BD Pyxis QFill Replenishment Station, REF: 155197-01, medication cabinet systemDue to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.
- Class II
Medical Device Recall · July 9, 2024
Medtronic DLP Single Stage Venous Cannulae with right angle metal tip, REF 67312During manufacturing, DLP Single Stage Venous Cannulae - straight tip (Model 66118) was incorrectly placed into a product labeled as DLP Single Stage Venous Cannulae - right angle metal tip (Model 67312).
- Class II
Medical Device Recall · July 9, 2024
BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900;There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care.
- Class II
Medical Device Recall · July 9, 2024
Covideien McGRATH MAC Video Laryngoscope, REF: 301-000-000, Non-Sterile, RxOnlyIFU update to address device's battery handling information.
- Class II
Medical Device Recall · July 9, 2024
Aurora Series, Aurora 7 lighthead; Models: LED7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
- Class II
Medical Device Recall · July 9, 2024
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and sSpecific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
- Class II
Medical Device Recall · July 8, 2024
IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versionsSafety Parameters Verification Mechanisms can be deactivated in clinical runtime
- Class II
Medical Device Recall · July 8, 2024
GM60A. Digital Diagnostic Mobile X-Ray System.Potential for the anti-fall system of the device arm to fail.
- Class II
Medical Device Recall · July 5, 2024
Medtronic ENT Ultra Round Steel Cutting Bur, Model Number REF 31313069Product was distributed past expiration date.
- Class III
Medical Device Recall · July 5, 2024
Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone iInvalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.
- Class II
Medical Device Recall · July 4, 2024
ORTHOPEDIATRICS CANADA ULC Dba PEGA MEDICAL
FRT250 - 4.0-6.4 CARTRIDGE. Component of the Fassier-Duval Telescopic IM SystemDuring the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of the FRT250-HEX.
- Class II
Medical Device Recall · July 3, 2024
GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Configured for Circular Staplers, 25mm; Catalog number: 1BSGC25;Devices were labeled with an expiration date of four years and ten days rather than the validated two years.
- Class II
Medical Device Recall · July 3, 2024
Siemens Medical Solutions USA, Inc
Sensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizatiIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
- Class II
Medical Device Recall · July 3, 2024
Siemens Medical Solutions USA, Inc
Sensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as welIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
- Class II
Medical Device Recall · July 3, 2024
Siemens Medical Solutions USA, Inc
Sensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or eIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.