- Class II
Medical Device Recall · October 10, 2022
Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
- Class II
Medical Device Recall · September 16, 2022
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
- Class II
Medical Device Recall · August 12, 2022
SURGIMESH XB, Polymeric Surgical Mesh, X-Large Ellipse Barrier Mesh, Model/Catalog Number: Tintra E-3030Potential presence of residual adhesive material on the mesh surface.
- Class II
Medical Device Recall · August 12, 2022
SURGIMESH WN, Polymeric Surgical Mesh, 14cm x 15cm Flat Sheet, Model/Catalog Number: T1415Potential presence of residual adhesive material on the mesh surface.
- Class II
Medical Device Recall · August 12, 2022
SURGIMESH XB, Polymeric Surgical Mesh, 14m x 15cm Flat Sheet with Silicone Barrier, Model/Catalog Number: Tintra R-1415Potential presence of residual adhesive material on the mesh surface.
- Class II
Medical Device Recall · August 12, 2022
SURGIMESH WN, Polymeric Surgical Mesh, Large Flat Sheet, Model/Catalog Number: T3030Potential presence of residual adhesive material on the mesh surface.
- Class II
Medical Device Recall · July 27, 2022
Diasol Acid Concentrate REF G100325-10Dex 100Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.
- Class II
Medical Device Recall · July 14, 2022
Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.
- Class II
Medical Device Recall · July 11, 2022
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
- Class II
Medical Device Recall · April 29, 2022
Adult Cap-ONE Biteblock REF YG-227T The cap-ONE Biteblock is intended to sample CO2 and administer supplemental oxygen, for patients who can wear a 20mm bite block, during endoscopy type proceduresDue to products being shipped/distributed to customers after the products expiration date had past.
- Class II
Medical Device Recall · March 14, 2022
Augustine Temperature Management, LLC
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model # WC77R (Refurbished); used in conjunHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Class II
Medical Device Recall · March 14, 2022
Augustine Temperature Management, LLC
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V; 2) WC71V-DEMO; 3) WC71V-R; used in conjunction with Warming Blankets anHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Class II
Medical Device Recall · March 14, 2022
Augustine Temperature Management, LLC
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC71, 2) Model # WC71R (refurbished); used in conjunction with Warming BlanketsHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Class II
Medical Device Recall · March 14, 2022
Augustine Temperature Management, LLC
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC77V; 2) WC77V-DEMO; 3) WC77V-R; used in conjunction with Warming Blankets and MHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Class II
Medical Device Recall · March 14, 2022
Augustine Temperature Management, LLC
Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses.HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Class II
Medical Device Recall · February 11, 2022
CombiDiagnost R90 1.0 (709030)While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
- Class II
Medical Device Recall · February 11, 2022
CombiDiagnost R90 1.1 (709031)While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
- Class II
Medical Device Recall · January 6, 2022
Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation systemMagnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
- Class II
Medical Device Recall · January 6, 2022
Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation systemMagnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
- Class II
Medical Device Recall · January 6, 2022
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation systemMagnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
- Class II
Medical Device Recall · October 28, 2021
Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device iIt was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
- Class II
Medical Device Recall · October 14, 2021
MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
- Class II
Medical Device Recall · July 15, 2021
(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
- Class II
Medical Device Recall · June 21, 2021
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rateDuring the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
- Class II
Medical Device Recall · March 1, 2021
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
- Class II
Medical Device Recall · December 24, 2020
PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce cross-sectional images of the body Model Number: 728143Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured
- Class II
Medical Device Recall · February 21, 2018
True 3D Viewer, Model: EP-00003With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.
- Class II
Medical Device Recall · April 11, 2016
TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinicianImaging system data may be displayed with an incorrect orientation (sidedness) where the patient side is not correct (left, right, superior, inferior, anterior or posterior). This may lead to an inaccurate measurements.
- Class I
Medical Device Recall · January 17, 2011
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, PRetrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.