Medical Device Recalls
RSS ↗7,979 medical device recalls from federal enforcement feeds.
Showing 7,701–7,750 of 7,979
- Class II
Medical Device Recall · November 28, 2023
MYNXGRIP 5F, REF MX5021, Vascular Closure Device (VCD). Used to seal femoral arterial and femoral venous access sites.Cordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.
- Class II
Medical Device Recall · November 28, 2023
EDS 3 CSF External Drainage System, REF: NT821731C, STERILEEO, Rx onlySterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.
- Class II
Medical Device Recall · November 28, 2023
Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUSTheir is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
- Class II
Medical Device Recall · November 28, 2023
Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
- Class II
Medical Device Recall · November 28, 2023
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
- Class II
Medical Device Recall · November 28, 2023
MYNXGRIP 6F/7F, REF MX6721, Vascular Closure Device (VCD). Used to seal femoral arterial and femoral venous access sitesCordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.
- Class II
Medical Device Recall · November 28, 2023
Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chairIn rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.
- Class II
Medical Device Recall · November 28, 2023
(1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
- Class II
Medical Device Recall · November 28, 2023
Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUSTheir is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
- Class II
Medical Device Recall · November 28, 2023
Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
- Class II
Medical Device Recall · November 28, 2023
VADER pedicle system, torque wrench, Catalog Number 42-702icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
- Class II
Medical Device Recall · November 28, 2023
(1) Patient Information Center iX, Model No. 866389 (2) Patient Information Center iX Expand, Model No. 866390 (3) PIC iX Hardware, Model No. 866424Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
- Class II
Medical Device Recall · November 28, 2023
External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlySterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.
- Class II
Medical Device Recall · November 28, 2023
VADER pedicle system, torque wrench, Catalog Number 42-703icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
- Class II
Medical Device Recall · November 27, 2023
Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100XThe following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
- Class II
Medical Device Recall · November 27, 2023
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems.A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.
- Class II
Medical Device Recall · November 27, 2023
RibLoc U Plus 90 Instrument PRIMARY GUIDE ASSY, LOW PROFILE REF RBL2320 The RBL2320, Low Profile Primary Guide is an instrument used with a power system to compress the RibLoc plates onto the boneDue to potential breakage during use.
- Class II
Medical Device Recall · November 27, 2023
The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY Torch Systems.A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.
- Class II
Medical Device Recall · November 27, 2023
Ortho-Clinical Diagnostics, Inc.
VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodiagnostic Products Anti-HBs Calibrators, Catalog No. 6801926Affected lots may experience increased calibration failures or an increase in falsely elevated results.
- Class II
Medical Device Recall · November 27, 2023
Mirion Technologies (Capintec), Inc.
Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures. Models: 5430-30151, 5430-30152,Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
- Class II
Medical Device Recall · November 23, 2023
Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membraneInterference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
- Class II
Medical Device Recall · November 23, 2023
FUJIFILM Healthcare Americas Corporation
The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
- Class II
Medical Device Recall · November 22, 2023
DeRoyal CARDIAC CATH SET-UP KIT PGYBK, REF 89-10572Kit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences.
- Class II
Medical Device Recall · November 21, 2023
DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.
- Class II
Medical Device Recall · November 21, 2023
Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.
- Class II
Medical Device Recall · November 21, 2023
Ortho-Clinical Diagnostics, Inc.
QMS Tacrolimus Calibrators, Product code 10015573Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.
- Class I
Medical Device Recall · November 20, 2023
Regard Dressing change kit, Item Number 800026006ROi CPS manufactured and distributed medical convenience kits with an irrigation solution component which was subsequently recalled by the component manufacturer.
- Class II
Medical Device Recall · November 20, 2023
Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
- Class II
Medical Device Recall · November 20, 2023
Calibration Serum Level 3 CAL2351CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
- Class II
Medical Device Recall · November 20, 2023
Olympus Corporation of the Americas
OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (5.4 mm diameter) Model: LTF-S190-5 Used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgicalInstructions for Use-User Deviations related to the reprocessing of the videoscope detailed in the Reprocessing Manual Instructions. The steps are required for the proper reprocessing of these videoscopes
- Class II
Medical Device Recall · November 20, 2023
Protege GPS Self-Expanding Peripheral Stent System, product number SERB65-09-60-80, Sterile.Seal defects could compromise the ability of the product packaging to maintain sterility.
- Class II
Medical Device Recall · November 20, 2023
LEICA BIOSYSTEMS NUSSLOCH GMBH
Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
- Class II
Medical Device Recall · November 20, 2023
KeySplint Hard Clear - Indicated for the fabrication of orthodontic and dental appliances such as mouthguards, nightguards, splints, repositioners, and retainers Item No.: 4220011Keyprint Keysplint Hard Clear 1 kg resin bottles were inadvertently packaged with the wrong Instructions for Use. Keysplint Hard Clear were placed inside a zip lock bag along with KeySplint Soft Instructions for Use (IFU).
- Class II
Medical Device Recall · November 20, 2023
NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3) Product Code S2190 - XLOne of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
- Class II
Medical Device Recall · November 20, 2023
Olympus Corporation of the Americas
OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (10 mm diameter) Model: LTF-S190-10 Used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical uInstructions for Use-User Deviations related to the reprocessing of the videoscope detailed in the Reprocessing Manual Instructions. The steps are required for the proper reprocessing of these videoscopes
- Class II
Medical Device Recall · November 17, 2023
Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
- Class II
Medical Device Recall · November 16, 2023
Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZDDue to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
- Class II
Medical Device Recall · November 16, 2023
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
- Class II
Medical Device Recall · November 16, 2023
V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
- Class II
Medical Device Recall · November 16, 2023
AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair proceduresPackaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE Kits, trays, and packs labeled as follows: CATHETERIZATION KIT, REF DYNDA2939Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
Centurion Kits, trays, and packs labeled as follows: a) DIALYSIS ACCESSORY PACK W/O CATHETER, REF MNS10750A; b) DIALYSIS ON KIT, REF DT21235A; c) INFUSA PORT KIT, REF DYNDC3166AMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE Kits, trays, and packs labeled as follows: a) ASC MINOR THYROID CDS, REF CDS983682F; b) BRONCHOSCOPY KIT, REF DYKE1718A; c) BRONCHOSCOPY KIT, REF DYKE1718B; d) DOUBLE CENTRAL LINE CAPMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
Medline Kits, trays, and packs labeled as follows: a) 709E RECOVERY PACK A,B,C II SS, REF DYNJ55960G; b) ABDOMEN PACK, REF DYNJ53643C; c) BAG TRUNK STOCK KIT, REF DYKM1880; d) BEDSIDE BRONCHMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
Centurion Kits, trays, and packs labeled as follows: MAINTENANCE FLUSH KIT/BLOOD DRAW, REF DT19335AMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class II
Medical Device Recall · November 15, 2023
Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE TRAY care and cleaning kits and trays for Tracheostomy and Stomas, labeled as: a) STOMA & TRACH CLEANING KIT, REF DYNDJ1129; b) TRACH CARE TRAY, REF DYNDJ1057A; c) TRACHEOSTOMY CARE KIT,Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
Centurion Kits, trays, and packs labeled as follows: a) A C DAILY WET KIT, REF DM405A; b) BWH DAILY DRIVELINE MGMT KIT, REF DM640; c) DAILY DRIVELINE MGMT SYSTEM, REF DM600; d) DAILY MAINTENAMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE Kits, trays, and packs labeled as follows: a) ARTERIAL LINE PACK, REF DYNJ65208A; b) CANCER SERVICE PORT KIT, REF DT22710; c) CAP CHANGE KIT, REF DYNDC2002A; d) CAP CHANGE KIT, REF DYMedline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- Class I
Medical Device Recall · November 15, 2023
MEDLINE INDUSTRIES, LP - Northfield
Medline Kits, trays, and packs labeled as follows: a) ANTERIOR PACK, REF DYNJ64814A; b) ARTHROGRAM TRAY, REF DYNDH1629; c) CVL INSERTION KIT-LF, REF DYNJ48587A; d) EXTREMITY CDS, REF CDS98331Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.