Medical Device Recalls
RSS ↗8,689 medical device recalls from federal enforcement feeds.
Showing 4,201–4,250 of 8,689
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.5T Power/Pulsar, Model Number: 781105;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
Welch Allyn CONNEX Accessory Power Management Stand:There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
- Class II
Medical Device Recall · March 24, 2025
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, UncementedThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
- Class II
Medical Device Recall · March 24, 2025
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, UncementedThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
- Class II
Medical Device Recall · March 24, 2025
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, UncementedThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.5T R11, Model Number: 781170;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.5T Explorer/Nova Dual, Model Number: 781108;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.5T Master/Nova, Model Number: 781106;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.5T, Model Numbers: 781195 and 781295;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Intera 1.0T Power/Pulsar, Model Number: 781103;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, UncementedThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
- Class II
Medical Device Recall · March 24, 2025
IceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- Class II
Medical Device Recall · March 24, 2025
Welch Allyn CP150 Electrocardiograph:There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
- Class II
Medical Device Recall · March 24, 2025
Ortho-Clinical Diagnostics, Inc.
Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains matIt was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
SmartPath to dStream for 3.0T, Model Number: 782145;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 24, 2025
IceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing applicationBoston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- Class II
Medical Device Recall · March 24, 2025
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
SmartPath to dStream for 1.5T, Model Number: 782146;The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Class II
Medical Device Recall · March 21, 2025
OEC 3DOEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn 2-Piece Reusable Blood Pressure Cuff Kits labeled as: 1. WELCH ALLYN 2-Piece Reusable Blood Pressure Cuff Kits, Product Code/Part Numbers: 2. WELCH ALLYN Reusable Blood Pressure CuffUndeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automateABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventOEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn Spot Vital Signs 4400 Device, Product Code/Part Numbers: 1) 44WT-2; 2) 44WT-3; 3) 44WT-4; 4) 44WT-6; 5) 44WT-B; 6) 44XT-2; 7) 44XT-4; 8) 44XT-6; 9) 44XT-B.Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn Connex Vital Signs Monitor, Product Code/Part Numbers: 1) ) Product Code/Part # 67MXTP-B-ECG3A; 2) ) Product Code/Part # 67MXTX-B; 3) ) Product Code/Part # 67MXXP-B; 4) ) ProductUndeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systemsCertain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn Welch Allyn Connex Spot Monitor, Product Code/ Part Numbers: 1) 73WT-B; 2) 74CE-B; 3) 74CT-B; 4) 74CX-B; 5) 74ME-B; 6) 74MT-B; 7) 74MX-B; 8) 74RE-B; 9) 74RT-B; 10) 75CUndeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn Green Series 777 Wall System, Product Code/Part Numbers: 1) 77791-2MP2X; 2) 77791-2MP2XL; 3) 77791-2MPX; 4) 77791-2MPX-HS; 5) 77791-2MPXL; 6) 77796-2MPX; 7) 77796-2MPXL.Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class I
Medical Device Recall · March 21, 2025
GE Medical Systems China Co., Ltd.
GE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn DuraShock Aneroid Gauge Sets with a Reusable Blood Pressure Cuff labeled as: 1) WELCH ALLYN DURASHOCK Aneroid Gauge Sets with a Reusable Blood Pressure Cuff, Product Code Part NumbersUndeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class II
Medical Device Recall · March 21, 2025
Welch Allyn Connex Integrated Wall System, Product Code/Part Numbers: 1) 84MTVE2-4; 2) 84MTVE2-6; 3) 84MTVE2-B; 4) 84MTVE2-US; 5) 84MTVEC-6; 6) 84MTVEC-B; 7) 84MTVEP-B; 8) 84MTVEX-4;Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
- Class II
Medical Device Recall · March 20, 2025
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope,There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
- Class II
Medical Device Recall · March 20, 2025
GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.
- Class II
Medical Device Recall · March 20, 2025
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablatioDue to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
- Class II
Medical Device Recall · March 20, 2025
DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemBeckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.
- Class II
Medical Device Recall · March 20, 2025
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
- Class I
Medical Device Recall · March 19, 2025
Brand Name: Ballard Product Name: Closed Suction Catheters Model/Catalog Number: 2271418-5 Software Version: N/A Product Description: Ballard Turbo-Cleaning Closed Suction System for Adults, 14 F,Lack of sterility assurance for closed suction catheter systems
- Class I
Medical Device Recall · March 19, 2025
Brand Name: Ballard Product Name: Closed Suction Catheters Model/Catalog Number: 220135 Software Version: N/A Product Description: Ballard Closed Suction System for Adults, 14 F, T-Piece ComponenLack of sterility assurance for closed suction catheter systems