Medical Device Recalls
RSS ↗7,959 medical device recalls from federal enforcement feeds.
Showing 4,701–4,750 of 7,959
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) MAJOR VASCULAR-LF, REF CDS840232L; 2) AVF-LF, REF CDS840233I; 3) GENERAL AAA #11-RF, REF CDS840261AB; 4) HEART CDS, REF CDS840387T; 5) OPEN HEART CDS PARTMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) BURN, REF CDS983465M; 2) BURN, REF CDS983465N; 3) BURN PACK, REF DYNJ63331DMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) TOT/ GYN LITHOTOMY CDS, REF CDS985431J; 2) TOT/ GYN LITHOTOMY CDS, REF CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, REF CDS985431L; 4) GENERAL LITHOTOMY PACKMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) MAJOR LITHOTOMY CDS-LF, REF CDS760056G; 2) LAVH CDS, REF CDS983411J; 3) MAJOR VAGINAL, REF CDS983497L; 4) DAY SURG LARGE VAG CDS, REF CDS984995N; 5)Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) CVL PACK, REF DYNJ00281O; 2) VASCULAR ACCESS PACK-LF, REF DYNJ0101291F; 3) MAJOR VASCULAR PACK-LF, REF DYNJ0488546K; 4) PEDS CENTRAL LINE PACK, REF DYNMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) ORAL SURGERY PK RFID, REF DYNJ52404I; 2) ORL PACK, REF DYNJ59073B; 3) ORAL MAXILOFACIAL PK, REF DYNJ80226A; 4) DENTAL MINOR PACK, REF DYNJ83140Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) CLEAN CLOSING PACK, REF DYNJ80686A; 2) CLEAN CLOSURE PACK GREEN, REF DYNJ84667Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) VAGINAL, REF CDS984955M; 2) DAY SURG LARGE VAG CDS, REF CDS984995M; 3) MAJOR VAGINAL PACK-LF, REF DYNJ0826768P; 4) VAGINAL PACK, REF DYNJ59081C; 5)VAMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) LITHOTOMY/PELVISCOPY PACK, REF DYNJ0531985R; 2) PELVISCOPY PACK-LF, REF DYNJ0843760I; 3) PELVISCOPY PACK-LF, REF DYNJ0843760JMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) ZALE HEAD & NECK CDS, REF CDS983184I; 2) ENT MINOR, REF CDS983466L; 3) EAR, REF CDS983486K; 4) HEAD AND NECK, REF CDS984259N; 5) HEAD AND NECK, REF CDS984Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 9, 2024
CORMIX 500 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64050; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.A limited number of IV bags have been found to leak during filling.
- Class II
Medical Device Recall · October 9, 2024
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) MAMMOPLASTY-LF, REF CDS983478L; 2) MAMMOPLASTY-LF, REF CDS983478M; 3) PLASTICS FREE FLAP, REF CDS983493M; 4) PLASTIC TORSO-LF, REF CDS983496L; 5) PLASTICMedline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
- Class II
Medical Device Recall · October 8, 2024
Approach CTO Micro Wire Guide, Guide Wire, Catalog Prefix CMW; Reference Part Numbers: CMW-14-190-12G CMW-14-190-18G CMW-14-190-25G CMW-14-190-6G CMW-14-300-12G CMW-14-300-18G CMW-14-300-25GAffected device lots have labels that state the incorrect expiration dates.
- Class II
Medical Device Recall · October 8, 2024
Approach Hydro ST Micro Wire Guide; Catalog prefix HMW Reference Part Numbers: HMW-14-135-ST HMW-14-190-ST HMW-14-300-STAffected device lots have labels that state the incorrect expiration dates.
- Class II
Medical Device Recall · October 8, 2024
MONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular and neurological tissuDue a component mix-up in manufacturing, the affected products may contain incorrect needle type/size and suture length.
- Class I
Medical Device Recall · October 7, 2024
Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, INPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class II
Medical Device Recall · October 7, 2024
DRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361The user of the assay may either face a failure message during re-calibration of the Hybrid.XL instrument using the re-calibration R1 and R2 of the device or find controls out of specification (failed Control High) which causes a delay in results.
- Class II
Medical Device Recall · October 7, 2024
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Philips Zenition 70, Model Number: 718133Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
- Class I
Medical Device Recall · October 7, 2024
Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GBPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class I
Medical Device Recall · October 7, 2024
LifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency tranPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class II
Medical Device Recall · October 7, 2024
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Philips Zenition 50, Model Number: 718096Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
- Class I
Medical Device Recall · October 7, 2024
Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, andPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class I
Medical Device Recall · October 7, 2024
Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KRPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class I
Medical Device Recall · October 7, 2024
Aeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transpoPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class I
Medical Device Recall · October 7, 2024
Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transpPotential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Class II
Medical Device Recall · October 5, 2024
CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces wiAs a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
- Class II
Medical Device Recall · October 5, 2024
CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces wAs a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4721Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4706Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
MEDLINE INDUSTRIES, LP - Northfield
CARROLLTON HIP FX Convenience kit used for surgical procedures, SKU SYNJ910003AProducts within a Medline pack may be impacted by a packaging error in which an incomplete seal has formed on the packaging thereby creating a packaging seal gap.
- Class II
Medical Device Recall · October 4, 2024
CAIRE FreeStyle Comfort portable oxygen concentrator, model #AS200-1.Potential failure of affected units to recharge the battery of the oxygen concentrators.
- Class II
Medical Device Recall · October 4, 2024
CAIRE FreeStyle Comfort portable oxygen concentrator, model #AS200-101.Potential failure of affected units to recharge the battery of the oxygen concentrators.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generateThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
MEDLINE INDUSTRIES, LP - Northfield
ARTHROSCOPY Convenience kit used for surgical procedures, SKU SYNJ910004CProducts within a Medline pack may be impacted by a packaging error in which an incomplete seal has formed on the packaging thereby creating a packaging seal gap.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4702Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke genThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
Medical Device Recall · October 4, 2024
CAIRE FreeStyle Comfort portable oxygen concentrator, model #AS200-2.Potential failure of affected units to recharge the battery of the oxygen concentrators.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke geneThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class I
Medical Device Recall · October 4, 2024
microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedureA complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4720Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4718Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by elThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated bThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
Medical Device Recall · October 4, 2024
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4716Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Class II
Medical Device Recall · October 4, 2024
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generaThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Class II
Medical Device Recall · October 4, 2024
CAIRE FreeStyle Comfort portable oxygen concentrator, model #AS200-103.Potential failure of affected units to recharge the battery of the oxygen concentrators.
- Class II
Medical Device Recall · October 3, 2024
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment ItemError code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay