Medical Device Recall · July 3, 2024
Ortho-Clinical Diagnostics, Inc.
VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.
- Class II
Medical Device Recall · July 3, 2024
Siemens Medical Solutions USA, Inc
Sensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post Processing Workstation-Intended to be used as a diagnostic and administrative tool supporting hemodynamic cathetIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
- Class II
Medical Device Recall · July 2, 2024
Siemens Medical Solutions USA, Inc
LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- Class II
Medical Device Recall · July 2, 2024
BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
- Class II
Medical Device Recall · July 2, 2024
CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 00811400218 0081Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.
- Class II
Medical Device Recall · July 2, 2024
Siemens Medical Solutions USA, Inc
Luminos dRF Max- Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762471The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- Class II
Medical Device Recall · July 2, 2024
Siemens Medical Solutions USA, Inc
Multitom Rax -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 10860700The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- Class II
Medical Device Recall · July 2, 2024
Stryker Sustainability Solutions
Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5ADisposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing, but that may have been reprocessed, which may lead to devices not performing as intended.
- Class II
Medical Device Recall · July 2, 2024
Siemens Medical Solutions USA, Inc
Luminos Agile Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762472The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
- Class II
Medical Device Recall · July 1, 2024
Neurovision Medical Products Inc
Cobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEOmislabeling; EMG RLN monitoring kit labeling contains the incorrect tube size.
- Class II
Medical Device Recall · July 1, 2024
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A BXB051502A BXB051901A BXB051902A BXB052901A BXB052902A BXB053901A BXB0Due to packaging defects, sterility assurance and heparin activity may be compromised.
- Class II
Medical Device Recall · July 1, 2024
Olympus Corporation of the Americas
The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgicalFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
- Class II
Medical Device Recall · July 1, 2024
Olympus Corporation of the Americas
The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgicalFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
- Class II
Medical Device Recall · July 1, 2024
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB055902C, BXB063902C, BXB065902C, BXB067902C, BXB073902C, BXB075902C, BXB077902CDue to packaging defects, sterility assurance and heparin activity may be compromised.
- Class II
Medical Device Recall · July 1, 2024
Olympus Corporation of the Americas
The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgicalFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
- Class II
Medical Device Recall · June 28, 2024
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.
- Class II
Medical Device Recall · June 28, 2024
IntelliVue Patient Monitor MX450 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866062. Indicated for use by health care professionals for monitoring and recording of, and toPhilips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
- Class II
Medical Device Recall · June 28, 2024
IntelliVue Patient Monitor MX550 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866066. Indicated for use by health care professionals for monitoring and recording of, and to gPhilips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
- Class II
Medical Device Recall · June 28, 2024
IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health care professionals for monitoring and recording of, and to gPhilips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
- Class II
Medical Device Recall · June 28, 2024
IntelliVue Patient Monitor MX400 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866060. Indicated for use by health care professionals for monitoring and recording of, and to gPhilips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack Number DYKE1558B; c) ENDO KIT, PaDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) LAMINECTOMY CDS, Pack Number CDS981510L; b) ORTHO TRAUMA-LF, Pack Number DYNJ900720D; c) ORTHO TRAUMA-LF, PackDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE1355; b) ENDO BASIC KIT, Pack Number DYKE1356; c) ENDOSCOPY KIT, Pack NumberDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVDTheir is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1092B; b) VIDEO BRONCHOSCOPY TRAY, Pack Number DYKE1666; c) SUCTION CANISTER KIT, PDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR PROCEDURE, Pack Number CDS981509C; b) GENERAL LAPAROSCOPY CDS, Pack Number CDS981512J; c) OR MAJOR CDS,Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline Semi-rigid suction liners labeled as follows: a) SEMI RIGID SUCTION LINER, REF OR515; b) SEMI RIGID SUCTION LINER WITH TUBING, REF OR526K; c) SEMI RIGID SUCTION LINER WITH TUBING, REF OR5Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: DENTAL, Pack Number DYNJ909451ADue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318260. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent pSome lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMINOPLASTY W/LIPO SDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
Capio SLIM Suture Capturing Device, 1-Pack, UPN: M0068318250. Capio SLIM Suture Capturing Devices are used by physicians to facilitate the consistent placement of sutures during open surgical proceduSome lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1350D; b) ENDO KIT, Pack Number DYKE1514BDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: KIT NEURO SHUNT ST LOUIS CHILD, Pack Number DYNJ907829BDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: MTHZ GYN LAPAROSCOPY, Pack Number DYNJ909739Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA0311C; b) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA0323C; c) ANESTHESDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: MERCY TIFFIN C-SECTION, Pack Number DYNJ909732Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic SIf temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) BAPTIST FLOYD BRONCHOSCOPY, Pack Number DYKE1456D; b) ENT MINOR, Pack Number DYNJ908099B; c) HEAD & NECK, PackDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: NICU ADMIT KIT, Pack Number DYKA1014GDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 27, 2024
MEDLINE INDUSTRIES, LP - Northfield
Medline convenience kits containing semi-rigid suction liners labeled as follows: ANES TURNOVER KIT-LF, Pack Number DYNJAA10338JDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JPCertain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
Siemens AG/Siemens Healthcare GmbH
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Class II
Medical Device Recall · June 26, 2024
ENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.